Olympus CelonLab ENT Інструкція - Сторінка 5

Переглянути онлайн або завантажити pdf Інструкція для Медичне обладнання Olympus CelonLab ENT. Olympus CelonLab ENT 10 сторінок. Safety check

Olympus CelonLab ENT Інструкція

2. Display function check

No.
Test accessories: Power cord
060
Connect the unit via power cord with a rated supply voltage. Switch on the unit and wait several seconds
until the power level display shows "1".
Press the "UP"-button of the power setting (increasing the power) ten times.
070
The power level display has to be increased by 1 per step until "10" (starting from "1") is shown.
Press the "UP"-button of the power setting (increasing the power) and hold it down.
080
The power level display has to be increased step-by-step until "25" is shown.
Press the "DOWN"-button of the power setting (reducing the power) ten times.
090
The power level display has to be reduced by 1 per step until "15" (starting from "25") is shown.
Press the "DOWN"-button of the power setting (reducing the power) and hold it down.
100
The power level display has to be reduced step-by-step until "1" is shown.
Documentation
Document the test results in "SCR.991007".
110

3. Power output

No.
Test equipment:
Test accessories: Power cord
120
Connect the unit via the power cord with a rated supply voltage. Connect the footswitch with the unit.
130
Connect the unit via the connection cables at the HF-output 1 and 2 (see fig. 3) with the electrosurgical
analyzer according to the analyzer's instructions for use and measure the following parameter:
Attention: Start measuring after the display shows the appropriate value and make sure to
release the footswitch before starting with a new measurement.
The power output at 5 (output load 100 Ω) has to be in the range of
140
The power output at 10 (output load 100 Ω) has to be in the range of
150
The power output at 15 (output load 100 Ω) has to be in the range of
160
The power output at 20 (output load 100 Ω) has to be in the range of
170
The power output at 25 (output load 100 Ω) has to be in the range of
180
Documentation
Document the test results in "SCR.991007".
190
© Celon AG
,
Issue date: 01.09.2008
SCI.991007-1.4
Safety Check
Instruction.doc
Safety Check Instruction
Electrosurgical analyzer (Example: QA-ES, Metron)
Connection cables, HF-output (banana) to the electrosurgical analyzer
Footswitch
Procedure
Procedure
4 W  P  6 W.
8 W  P  12 W.
12 W  P  18 W.
16 W  P  24 W.
20 W  P  30 W.