PolyLoop Ligating Device
Quick Reference Guide
The PolyLoop delivers a nylon loop snare designed to prevent or control bleeding following
polypectomy of pedunculated polyps.
CAUTION:
This guide is only a summary of the instructions for use for the HX-400U-30 PolyLoop. Be sure to follow the Olympus EndoTherapy
PolyLoop HX-400U-30 instruction manual that was included with your HX-400U-30 purchase.
1
Look at the polyp for signs of ligation (blanching
at the stalk, engorging in the head).
2
Close the snare at least 2 mm above the PolyLoop
on the stalk of the polyp.
1
Straighten the endoscope and
n
PolyLoop as much as possible.
(See images A and B.)
Push the slider distally to extend
n
the hook from the coil sheath.
(See image C.)
2
Cut the sheath, coil, and operating wire near
n
the yellow cylinder using pliers or a wire cutter.
(See image at right.)
Withdraw the endoscope.
n
3
Insert the scope next to the sheath. Use the
n
Loop Cutter to cut the PolyLoop between the
stopper and the coil sheath.
Remove the sheath from the patient.
n
Using an Electrosurgical Snare:
(See image at right.)
Releasing a Trapped PolyLoop Ligating Device:
A
(See image at right.)
3500 Corporate Parkway, PO Box 610, Center Valley, PA 18034
( HX-400U-30 )
Olympus is a registered trademark of Olympus Corporation, Olympus America Inc., and/or their affiliates.
PolyLoop
B
C
Loop stopper
Coil sheath
For more information, contact your local Olympus
sales representative, or call 800-848-9024.
www.olympusamerica.com
©2013 Olympus America Inc. All rights reserved.
Printed in the USA OAIET0613QRG11197