OLYMPUS CIX100 Cleanliness Inspector System
8.2 Mounting and inspection of tape lift samples in the CIX100 system
It is recommended to follow the instructions described in the ASTM E1216-11
standard reapproved in 2016 for tape lift inspection with the CIX100 system. The
recommended type of tape is 3M No. 480. It is not guaranteed that other types of
tapes can be used with the CIX100 system.
1. To mount the tape to the sample holder, place the frame on the holder such
that the markings 50 and 19 face upwards.
2. Glue the tape to the frame such that the 50 mm wide tape is inside the
respective markings. Make sure that the sample is as flat as possible and be
careful not to stretch the tape while mounting.