Olympus MX61A Посібник - Сторінка 18
Переглянути онлайн або завантажити pdf Посібник для Лабораторне обладнання Olympus MX61A. Olympus MX61A 48 сторінок. Motorized 300 mm compatible semiconductor/fpd inspection microscope
@
15
@
Fig. 8
@
Fig. 9
Fig. 10
3
Attaching the Objectives
# Never attempt to rotate the motorized revolving nosepiece by hand.
Rotating by hand may cause malfunction or damage. Remove the
cap from each objective mount hole.
1. Following the order of numbers 1 to 5 or 6 indicated on the objective
mount thread @, screw objectives into the objective mount threads by
starting with the objective with the lowest power.
# The rotation of the revolving nosepiece can be made smooth by
mounting objectives uniformly in all of the nosepiece positions.
Therefore, the objectives that are not used frequently should also be
mounted. Be sure to attach the caps to the unused objective mount
hole.
Using the Brightfield Objective Adapter
When mounting a brightfield objective on a revolving nosepiece designed
for brightfield/darkfield objectives, mount each objective by means of the
BD-M-AD objective adapter.
Note that this adapter degrades the autofocusing performance if it is
used during autofocusing (using a high-power objective).
4
Attaching the Motorized Revolving Nosepiece
# It is only the U-D6BDREMC, U-D5BDREMC, U-D6REMC or U-P5BDREMC
revolving nosepiece that can be used.
# When mounting the revolving nosepiece, remove the stage holders
to ensure that the objectives do not touch the stage.
# Slide the revolving nosepiece all the way into the dovetail mount
until it stops.
# The revolving nosepiece is a heavy module. Take care not to drop it
during mounting.
1. Loosen the revolving nosepiece clamping screw @ using the provided
Allen screwdriver, slide the revolving nosepiece all the way into the dove-
tail mount until it stops, and tighten the clamping screw @.
# If the clamping screw is tightened without sliding the revolving nose-
piece all the way, the clamping section may be deformed, the re-
volving nosepiece may become irremovable, and the connector
contact failure may occur also.
2. Set the DIP switches for the mounted revolving nosepiece (see page 22
for the setup method).
5
Attaching the Observation Tube
# Do not hold the observation tube by the binocular section. Since the
binocular section is movable, the observation tube may not be fitted
into the circular dovetail correctly resulting in incomplete attachment.
1. Using the provided Allen screwdriver, fully loosen the observation tube
clamping screw @.
2. Insert the circular dovetail mount at the bottom of the observation tube
into the observation tune mount on the top surface of the arm.
3. Clamp the observation tube by tightening the clamping screw @.
(Fig. 8)
(Fig. 9)
(Fig. 10)