Olympus MX61A Посібник - Сторінка 38

Переглянути онлайн або завантажити pdf Посібник для Лабораторне обладнання Olympus MX61A. Olympus MX61A 48 сторінок. Motorized 300 mm compatible semiconductor/fpd inspection microscope

CONFIG 5/6 screen
35
@
7
³
|
...
²
ƒ

CONFIG 5/6 Screen Setup

}The CONFIG 5/6 screen is used to fine tune the aperture iris diaphragm
(AS) values.
}If the AS preset values set in the CONFIG 1/6 screen cannot be varied,
touch the NEXT switch.
}If "- - -" was set in the CONFIG 1/6 screen, the fine tuning value cannot be
set. (In this case, the AS SET switch ³ is grayed out.)
1. If it is required to modify a default value, touch the desired objective No.
switch @ for the objective to be fine adjusted.
2. Press the AS open or close switch ² a few times to adjust to the desired
fine tuning value.
3. Touch the AS SET switch ³ to enter the setup. The AS FINE TUNING field
of the objective shows the difference from the AS PRESET value.
4. Set the fine tuning values of other objectives by repeating steps 1 to 3 for
each objective.
Other Switches
|AS RESET: Touch this switch to reset the setting.
ƒBACK: Touch this switch to return to the CONFIG 4/6 screen without
applying the fine tuning value setup.
...NEXT: Touch this switch to advance to the CONFIG 6/6 screen by
applying the AS fine tuning values.