6
Pneumatic and Power
6.1
Pneumatic
1. Connect 12mm CDA input tubing to the machine. Make sure the pneumatic disconnect is open to
allow air flow. Refer to figures 20 and 21.
2. The regulators have been set to 0.5 MPa at the Axometrics facility. However, if pressure
adjustments need to be made, raise the regulator knob and turn CW or CCW to raise or lower the
pressure. Make sure to lower and "lock" the adjustment knob once the digital pressure gauge
reads 0.5 MPa. Refer to figure 22.
3. Check to make sure the granite is floating in range on the scale. Refer to figure 23.
Fig.20
Air Input
Regulator Adjustment Knob
Fig. 22
Isolators Regulator
Fig. 23
Digital Pressure Gauge
12
Fig.21
Pneumatic Disconnect in Open Position