a) ATC mode: Put the selection knob on "℃". The value displayed means the temperature of the
medium to be measured. At the same time the temperature signal be sent to the pH-t mixture circuit
to compensate the pH value.
b) MTC mode: Remove the temperature sensor, Put the selection knob on MTC and "℃" position.
Adjust the temperature regulation knob until the value of temperature displayed is same as the
indication value of thermometer pluged in the medium to be measured. At the same time the
temperature signal be sent to the pH-t mixture circuit to compensate the pH value manualy.
2.3 Calibration
Before operating, i.e. before measuring the solution to be measured, the instrument should be
calibrated. However, it does not mean that it must be calibrated before operating each time. For
successive operating, in general, it will be calibrated once a day.
There are two methods of calibration, one is one point calibration for normal use and the other is two
points calibration for exact measuring.
2.3.1 One point calibration procedure (taking ATC for example):
a) Remove the Q9 short circuit plug, Link the compound electrode and the temperature sensor, clean
the electrode with distilled water and blot it with filter paper, then plug the electrode into the buffer
solution (pH=4 or pH=9.18).To measure the temperature first, put the selection knob on "℃" and
ATC position, The value displayed means the temperature of the medium to be measured.
b) Put the selection knob on "pH", turn the slope knob to the end clockwise, adjust setting regulator
until the value displayed equals to the standard pH (pH=4 or pH=9.18).
Note: The "setting" knob will not be changed after calibration finished.
2.3.2 Two points calibration procedure:
a) Taking ATC for example. Remove the Q9 short circuit plug, Link the compound electrode and the
temperature sensor, clean the electrode with distilled water and blot it with filter paper, then plug the
electrode into the buffer solution (pH=6.86).To measure the temperature first, put the selection knob
on "pH", The value displayed means the pH of the standard buffer solution (pH=7).
b) If the solution to be measured is acidic, take out the electrode from the solution of known pH value
(pH value =6.86), clean the electrode with distilled water and blot it with filter paper, then plug it into
the buffer solution of known pH value (pH value = 4). Adjust the slope regulator until the value
displayed means the pH of the standard buffer solution (pH=4).
c) If the solution to be measured is alkalescent, we can calibrate the meter secondly with the solution
of known pH value (pH value =9.18). Adjust the slope regulator until the value displayed means the
pH of the standard buffer solution (pH=9.18).
The calibration of the instrument is completed. Once the instrument has been calibrated, its SETTING
potentiometer should not be changed any more .
2.3.3 Calibration with MTC to reference 2.2 (b).
3. pH value measurement
a) In case that the temperature of the solution to be measured the same as the temperature of the
setting solution.
Rinse the end of the electrode with distilled water, and suck it dry with clean filtering paper.Put the