7. Advanced

This section contains instructions for advanced features of the C-Test station:
Accessing the detector bump logs and event logs
Configuration programming for both the monitors and C-Test
Firmware programming for both the monitors and C-Test
7.1.

Functionality

The table below details the functionality available
C-Test Manager Functionality
Retrieve event logs and bump test logs
Program configuration files for both monitors
and C-Test
Upload Firmware Upgrades for both monitors
and C-Test
For more information about C-Test Manager, please refer to
7.2.

Retrieving data

The section below will cover retrieving data manually by removing the USB from the C-Test and
inserting it into a computer:
Once you have manually accessed the USB through a computer you will see up to 3 folders: Logs,
Firmware, and Config. When accessing the USB directly only the Logs folder will apply, use caution
when accessing other folders as they contain files that can only be changed through the C-Test
Manager Software.
Caution: Do not delete any files from the USB prior to saving them; the C-Test does not store this
information internally outside of the USB.
7.3.

Logs Folder

The C-Test is designed to automatically collect the results of each bump test for each monitor. In
addition, it will download the detector event log. All detector event logs will be stored in the folder
labelled 'CC' and all test logs will be stored in the 'TS' folder. Each time a detector is tested it will
transmit the entire event log up to 25 events to the USB memory. These events are saved as individual
logs. It will be common for a monitor to have multiple event log files stored on the USB. Individual event
logs will be organized inside the Clip folder by the serial number. The serial number format is in the
form of PGYY123456 (where P=Product code, G=gas type, YY=year, and 123456 is the 6-digit serial
number). The logs will be stored in logs / CC / PGYY / (serial / 512) / [fullserial#]
Every bump test and calibration performed on the C-Test is stored into the file named TS-CC.csv,
inside the TS-CC folder. This file follows a specific format so that it can later be parsed into a database
or Excel format for further manipulation and storage. You can then access the CSV file containing the
test results. There are multiple lines in every test report. We will illustrate how to read this report line by
line.