Dini Argeo TPW20IEX2GD Інструкція з техніки безпеки - Сторінка 4

Переглянути онлайн або завантажити pdf Інструкція з техніки безпеки для Навантажувачі Dini Argeo TPW20IEX2GD. Dini Argeo TPW20IEX2GD 18 сторінок.

 For weighing objects made up of light metals (aluminium, magnesium and titanium or their alloys, excluding the alloys
with less than 10% of aluminium, the varnishes and coatings with less than 25% of aluminium in weight), the use of
stainless steel forks is necessary.
 The use of the instrument in hazardous areas require a special attention and special precautions during the use and
maintenance
 The instrument conforms for use in zones having precise features: do not install and use the instrument in environments
different than those provided for.
 The installation, maintenance and repair of the instrument, must be made by qualified and authorized personnel and
with the approval of expert technicians.
 The safety of the explosion-proof system is guaranteed only if the system is installed, used and taken care of following
the instructions given in this manual and in the technical manual (TECH.MAN.REF.).
 Check that the pallet truck scale is in good operating conditions.
 Only spare parts approved by Dini Argeo should be used.
 Be very careful when using the instrument; any sparks could cause an explosion.
 Avoid accumulations of dust. To avoid dust layers, the user must regularly clean the surface of the enclosure,
considering that the maximum thickness allowed is < 5mm.
 WARNING! ELECTROSTATIC HAZARD, CLEAN ONLY WITH WET CLOTHS OR ANTISTATIC PRODUCT
 Avoid accumulations of electrostatic charges; therefore, when using the instrument in a hazardous zone, the appropriate
work clothing must be used by the operator or the maintenance person.
 Do not cover the instrument with coverings made by materials which could have electrostatic charge
 Do not paint the system
 It is forbidden to modify or repair the instrument with components not conforming to the certification; this action
compromises the intrinsic safety of the instrument (with a subsequent loss of the Ex approval) and the nullification of
the product warranty.
 It is forbidden to connect the instrument to modules not provided for by the certification; this action compromises the
intrinsic safety of the instrument (with a subsequent loss of the Ex approval). Contact Dini Argeo srl for further
information.
 Avoid direct sunlight.
 NEVER load the scale beyond the maximum capacity shown on the weight indicator's plate.
 Avoid the use of oxidising substances (hydrogen peroxide).
TPWX2GD/TPWX2GDI/TPW20IEX2GD

3. !! WARNINGS !!