Alcolock L Посібник з експлуатації - Сторінка 5

Переглянути онлайн або завантажити pdf Посібник з експлуатації для Датчики безпеки Alcolock L. Alcolock L 8 сторінок. Alcohol interlock

CALIBRATION PROCEDURES

It is recommended calibration be done indoors in a service facility, using an ACS
download station.
The calibration procedures are intended for use by trained technicians only.
EQUIPMENT NEEDED
• 130 PPM (0.05 g/dL) alcohol gas standard, ACS #95-000425 or 260 PPM
(0.10%BAC) alcohol gas standard, ACS #95-000436
• Regulator valve assembly, ACS #79-005546
• V3 download station, ACS #79-006111
• V3/L OEM download station adapter, ACS #79-009115
COMPRESSED GAS SAFETY
Calibration should be performed indoors in a service facility, where the gas
standard cylinder can be properly stored.
Transportation and storage of compressed gases in vehicles is dangerous and should be avoided.
Other precautions:
• Never remove or alter canister labels
• Remove the valve and install the protective cap on cylinders when not in use
• Store cylinders in a cool, well ventilated area, away from sources of heat
CYLINDER SETUP
1.
Examine the canister and valve for any damage, and check the expiration date on the label.
Observe all cautions and safety warnings found on the canister.
2.
Remove the protective cap and screw the regulator valve to the canister. Attach the rigid
delivery tube. Never modify the delivery tube in any way.
CAUTION! Damaged or broken valves can turn a canister into an unguided missile.
Attach the valve in a safe location.
SERVICE GUIDE
3