AlcoMate Core Посібник оператора - Сторінка 5
Переглянути онлайн або завантажити pdf Посібник оператора для Випробувальне обладнання AlcoMate Core. AlcoMate Core 8 сторінок. Digital personal breathalyzer
Warnings and Precautions
WAIT AT LEAST 20 MINUTES AFTER EATING OR DRINKING BEFORE TESTING. A breathalyzer
examines "lung air" samples, not "mouth air" to calculate alcohol content. Any "contaminants" in the
mouth, including alcohol, food, smoke, or even excess saliva, may interfere with testing. This
includes alcohol-based mouthwash or similar products. This is the most common cause of improper
use and incorrect results.
Avoid testing in high wind or enclosed spaces. Poor ventilation may also lengthen the time required
for warm up between tests.
Do not blow smoke, saliva or other contaminants into the mouthpiece, as the sensor may become
damaged.
Do NOT use the AlcoMate Core as a tool to drink and drive.
Keep out of extreme temperatures and out of the reach of children.
To maintain accuracy, the sensor inside the AlcoMate Core must be recalibrated every 200 tests or
at least once a year (whichever comes first). Large numbers of "high positive" (high alcohol)
readings may require recalibration of the sensor much sooner than normally expected. To
recalibrate, you must mail your device to the address at the end of this manual. Please contact us
for details and related fees.
Troubleshooting
WARNING DISPLAY "Flo": If you see "Flo" in the display, the breath sample was not strong or long
enough for measurement. Please try again and make sure the breath is strong and steady during
sampling. As soon as you start blowing into the device, you will hear a long beep indicating how long
you should blow.
WARNING DISPLAY "SEn": If you see "SEn" in the display, there is a general sensor module error.
Please contact us for technical service.
WARNING DISPLAY "bAt": If you see "bAt" in the display, the battery power is low. Please replace
the batteries with a brand new pair. Make sure to properly align the batteries by looking at the +/-
marks.
WARNING DISPLAY "Err9": Power source is too weak to complete operation. Please replace
batteries with a new pair.
For the other troubles, please contact AK GlobalTech at [email protected], along with your
error codes and photos if available.
- Warnings and Error Codes -
5