8.10 Taking the System Apart
The Control Unit is designed so almost all components
can be removed without disassembling other parts.
Open the housing by removing the 4 screws from
1
the bottom part.
Remove PCBA by disconnecting the cables and unscrewing
2
the 2 screws.
Remove the compressor by unscrewing the 4 screws
3
and disconnect the power cable from the PCBA and the
air tube on the compressor.
Remove the air out/power in part by disconnecting the air
4
tubes and the cables from the PCBA.
Remove the rotor valve by pulling out the locking pin
5
from the motor axes and removing the air tubes from
the rotor valve.
Remove the rotor valve motor by unscrewing the 2 screws.
6
Remove the micro switch by unscrewing the 2 screws.
7
Removing the hooks requires the compressors to be removed
8
first. Remove the four screws on the inside of the back part
of the housing and then release the attachment part for the
hooks from the outside of the housing.
8.11 Reassembling the System
To reasemble the Control Unit, reverse the steps in 8.10 ensuring
that the pressure sensor tubing runs underneath the compressor
during reassembly.