Daily Maintenance

5
5
Daily Maintenance
This chapter describes how to maintain the product.
To maintain the performance of this product, Nikon recommends you to perform daily maintenance.

Cleaning Utensils and Consumables

5.1
5.1
Cleaning utensils
(Note)
 Soft-tip brush
 Soft cloth (nonwoven cloth, cotton cloth, or gauze)
Note: Use the cleanroom wiper in the cleanroom.
Consumables
 Neutral detergent (only when the product main body is very dirty)
5.2

Cleaning the Camera Body

5.2
 To clean the product, see that the F-mount cap or the F-mount adapter is attached to it, and then wipe off stains on the
DS-Ri2/DS-Qi2 body using a soft cloth or others. If the product is very dirty, gently wipe stains using gauze moistened with
a small amount of diluted neutral detergent.
 Do not use organic solvent (such as alcohol, ether, and thinner) on coated parts, plastic parts, or printed parts. It causes
discoloration or exfoliation of printed characters.
 Blow off dust in the mount with an air blower.
Cleaning the mount interior
Blow off dust in the mount using an air blower. Do not touch the components as long as possible. Especially, never touch
the glass filter on the front of the image pickup device.
If the glass filter of the image pickup device is touched, or dust adhered to the glass filter cannot be removed, contact your
nearest Nikon representative for cleaning.

Cleaning Optical Devices

5.3
5.3
Prevent dust or fingerprints from being attached to the lens of optical devices. Stains on the lens degrade the visibility of
images.
For details on how to clean optical devices, see the instruction manual of your optical device.

Storage

5.4
5.4
 When the product is not in use, turn off the power. When the product is not used for a long period of time, unplug the
power cord of the AC adapter.
 Store the product in a location satisfying the storage environment conditions described in "Chapter 7 Major Specifications."
 Store this system in a location with low humidity where mold is unlikely to form.
 Make sure to attach the F-mount cap to store the DS-Ri2/DS-Qi2.
 During storage, place a plastic cover over the equipment to prevent dust accumulation.
Before placing the plastic cover, turn off the power switch of the product and wait for the product to cool down.

Periodic Inspections (Charged)

5.5
5.5
Periodic inspections are recommended in order to maintain the performance of this product. Consult your Nikon
representative for details about periodic inspections.