Fluorescence Observation

17

Warning

17.1
The mercury lamp (or xenon lamp) used with the Epi-illumination attachment or
DIH-M requires careful handling. Be sure to read the warnings described in the
beginning of this manual and in the operating manual provided by the manufacturers
of the super high-pressure mercury light source (or high-intensity light source) and
lamp. Observe all the warnings and precautions described in those documents.

Epi-illumination attachment shutter

17.2
The shutter blocks illumination. When suspending
microscopy, close the shutter to prevent fading of
specimen colors. (Set the shutter lever to the C
position to move the shutter into the optical path
and block light.) To protect important specimens,
make it a habit to use the shutter whenever
appropriate.
When pausing Epi-fluorescent microscopy to
perform microscopy diascopic light, move the
shutter into the optical path to block the
Epi-illumination light.
When using the Epi-illumination attachment:
Set the shutter open/close lever to the "C"
position to set the shutter into the optical path
and to block illumination.
When using the DIH-M:
Press the hand switch to toggle the shutter
open or closed. The shutter is automatically
kept closed when the DIH-M power switch is
turned on.

Light shielding plate

17.3
The light shielding plate protects the observer's eyes from ultraviolet light reflected from the
specimen. To remove the plate, loosen the clamp screw and pull it forward.
Individual Operations
17 Fluorescence Observation
26
Shutter lever
Light shielding plate
Hand switch