AMC-CD2 Portable CO2 Detector

5 PREVENTIVE MAINTENANCE

5.1 GENERAL

The enclosure should be brushed or wiped clean once a year or more, depending on the
accumulation rate of dust or dirt.
To avoid sensor damage, the unit MUST NOT be submerged in any liquids. Hosing or splashing
the unit with any liquids must also be avoided.

5.2 VERIFICATION OF OPERATION

The operation and response of the CD2 may be verified by exhaling near the sample inlet. The
display should show an elevated reading. This test should be performed as required to confirm
the CO2 sensitivity.

5.3 USER MAINTENANCE

The inline transparent filter should be replaced when the sample flow is impeded or when the
filter element appears dirty. The replacement filter part number is AMC-126. The sample hose is
durable, kink resistant and is of a self- resealing natural latex. To obtain a replacement hose,
order part number AMC-804R and specify the required length.
A variety of sample tips are available. Please refer to Section 2.3 Sampling Needles for the list
of tip sizes, types and part numbers.