DAILY LEAK
CHECK PROCEDURE
5
Press the oxygen flush button and hold until
the pressure on the manometer indicates
approximately 30 cm H
7
Caution:
Open the evacuation valve before using
the anesthesia machine again .
805-0975-030
O .
2
6
Check the manometer . The pressure must not
drop (But it can increase) . If the pressure drops,
replace the patient circuit and the bag and try
again .
8
If you encounter an excessive leak, it is
recommended that you proceed by elimination
to find the source of the problem by replacing a
single component at a time . The most frequent
leaks can be found at the fresh gas outlet o'ring,
the patient circuit and the bag .
Once the leak is identified, dispose of the
defective item to avoid it being reused by error .
At the end of each day, it is recommended that
you close the oxygen tank and fill the
anesthetic agent vaporizer .