EDI CTT-2 Посібник з експлуатації - Сторінка 3

Переглянути онлайн або завантажити pdf Посібник з експлуатації для Випробувальне обладнання EDI CTT-2. EDI CTT-2 6 сторінок.

2.1.3
is made. A short delay will speed up the sweep at the expense of accuracy. A delay of
1000uS is a good compromise between speed and accuracy.
2.2
A typical sweep test for a 200 KHz transducer would cover a range of 170 to 230
Khz using 100 Hz steps. The number of steps required will be the sweep width divided by
the frequency step. In this example, the frequency sweeps from 170 to 230 Khz for a width
of 60KHz. Using 100Hz steps this will require a total of 600 steps to cover the range. The
recommended delay should be left at 1000uS. The 'Sf' command sets the frequency step,
'Ss' sets the number of steps, and 'Sd' sets the step delay. Typing 'S' will display the current
frequency step, number of steps, and delay. For this example, type 'Ss600' to set the
sweep for 600 steps.
2.3
To start the sweep, type 'sm170000' and the sweep will start at 170KHz and
continue for 600 steps to 230KHz. When the sweep is finished the maximum and minimum
impedances found will be displayed along with the frequency that they ocurred at. If
impedance data for each frequency step is desired, use the command 'sa170000' and the
impedance and relative phase will be returned for each frequency step.
3 Spot frequency tests measure the impedance at a single frequency. To check the impedance of
the transducer at 197KHz, type 'f197000'. To read both impedance and the relative phase, type
'fa197000'. The 'f+' command will increment the frequency one step, and the 'f-' commad will
decrement the frequency one step. The impedance and relative phase are returned.
4 Leakage testing is done with the 'l' command, and returns the resistance of the transducer. The
leakage test takes 3 to 10 seconds, the longer time will read higher resistances. A minimum
delay of 3000mS is required. To set a 5 second delay, type 'Ld5000'. Note that the commands
are case sensitive.
5 The three indicator lights on the front panel can be turned on, off , or flashed to indicate what
action is to be taken during the test. To light the left most (green) LED, type 'I1', to extinguish
the LED, type 'i1'. The LED's are numbered from left to right 1, 2, and 3. The corresponding
colors are green, yellow, and orange. Typing 'I2' lights the yellow LED, and 'I3' lights the
orange one. To make a LED1 blink, type 'Ib1' and the green LED will begin flashing. To stop
the LED from flashing type 'i1'.
6 Returned data format
Two data formats are provided, one returns descriptive data, and the other returns only comma
separated (CSV) data. When first trying out the CTT232, is is easier to use the default
descriptive data that returns data in the form F= 100000, Z = 257 instead of the CSV that will
only return 100000,257. The 'p' command switches to CSV data. To go back to the descriptive
data, type 'P'.
CTT-2 Manual V1.01
The delay is the time that the frequency is held constant before the measurement
3