To calibrate the gauge, select CALIBRATE from the channel 1 or 2 menu
and press the ENTER key. Select FULL CAL and press ENTER. The display
will read SET LO POINT ON CALIBRATOR AND PRESS ENTER. Adjust the
calibrator/simulator at or near zero or a very low reading and press ENTER;
the display will show SAMPLING, then ADJUST LO POINT TO MATCH
CALIBRATOR. Use the or arrow keys to increase or decrease the display
reading to match the setting of the simulator and press ENTER. The display
will show SET HI POINT ON CALIBRATOR AND PRESS ENTER. Adjust the
simulator at or near the span value of the transducer or a very high reading
and press ENTER; the display will show SAMPLING, then ADJUST HI POINT
TO MATCH CALIBRATOR. Again use the or arrow keys to increase or
decrease the display reading to match the simulator and press ENTER. The
display will read CALIBRATION VALUES SAVED!. The gauge will return to the
home screen with the new calibration values stored in memory.
8.4.2
The CDM-2100 gauge has a feature that allows a slight adjustment of either
the zero or span values individually. This type of calibration can be used
to "tweak" the readout to match that of a known value without actually
performing a formal calibration procedure. This adjustment is independent for
each channel and must be performed on that individual channel. Please note
that this type of adjustment will invalidate calibration settings from the FULL
CAL procedures.
TWEAK LO ONLY
To make a small adjustment on the zero calibration value of the gauge, enter
the calibration mode by selecting CALIBRATE and press ENTER; select
TWEAK LO ONLY from the menu and press ENTER. The display will show
SET LO POINT ON CALIBRATOR AND PRESS ENTER. Adjust the calibrator/
simulator at or near zero or a very low reading and press ENTER; the display
will show SAMPLING, then ADJUST LO POINT TO MATCH CALIBRATOR.
Use the or arrow keys to increase or decrease the display reading to
match the calibrator and press ENTER. The display will read CALIBRATION
VALUES SAVED!. The gauge will return to the home screen with the new zero
calibration value stored in memory.
TWEAK HI ONLY
To make a small adjustment on the span calibration value of the gauge, enter
the calibration mode by selecting CALIBRATE and press ENTER; select
TWEAK HI ONLY from the menu and press ENTER. The display will show
SET HI POINT ON CALIBRATOR AND PRESS ENTER. Adjust the calibrator/
simulator at or near the desired span value and press ENTER; the display
will show SAMPLING, then ADJUST HI POINT TO MATCH CALIBRATOR.
Use the or arrow keys to increase or decrease the display reading to
match the calibrator and press ENTER. The display will read CALIBRATION
VALUES SAVED!. The gauge will return to the home screen with the new span
calibration value stored in memory.
8.4.3
RECALL FACTORY CAL VALUES
The user can at any time during the life of the gauge reinstate the factory
calibration values for channel 1 or 2 independently. Select CALIBRATE from
the CHANNEL 1 or CHANNEL 2 menu and press ENTER; select RECALL
FACTORY CAL and press ENTER. The next screen will display the type and
range for which the selected channel's input is currently configured. Select
APPLY to confirm or CANCEL to decline and press ENTER. If APPLY is
selected, the display will show CALIBRATION VALUES SAVED!. The gauge will
return to the home screen with the factory default calibration values stored in
memory. If CANCEL is selected, the gauge will retain the current calibration
values. Press the ESC key to return to the home screen.
The calibration values only, will return to the factory default; all other settings will
remain unchanged. If the transducer type or range is incorrect, press the MENU/
ESC key to abort saving incorrect factory cal values. Configure the gauge for the
desired input sensor type and range and then recall the factory cal values.
8.5
RESET
The reset selection in the menu is used to reset the min/max reading for
channel 1 or 2 independently. To perform a reset, select either channel 1 or 2