RECOMMENDED CLEANING PROCEDURE
Hettich Centrifuge – Rotors and Accessories
1. Clean with soap & water or a mild cleaning
agent after every use (or at least once a week)
2. If parts are contaminated by pathogenic or biologic
material we recommend:
– For disinfection: Helipur
B. Braun Melsungen
3. Dry immediately after cleaning
4. Lightly grease aluminium parts after drying using
acid-free grease, e.g. vaseline
Rotors and attachments may no longer be used upon indication of wear and tear or corrosion!
H plus N from
®
© Arthrex GmbH, 2016. All rights reserved. LI2-00016-EN_A
5. With aerosol-sealed rotors and bio safety systems:
– Clean & check sealing rings regularly (weekly)
– Replace sealing ring immediately upon indication of
crack formation, embrittlement or abrasive wear
– Lightly grease seal ring with acid-free grease,
e.g. vaseline
6. Rotor and Motor Shaft (clean once a month):
– Disassemble the rotor and clean
– Lightly grease the motor shaft
7. Checked all parts on a monthly basis for
corrosion damage