We are now going to remove the faulty load cell cabling from the Molex connector.
The four Load Cells connect into the XPV7 Electronics Module (PCB) and are colour coded on the
load cell cables at the terminal blocks for ease of wiring, they are as follows
Load Cell 1.
Blue
Heat - Shrink shroud
Load Cell 2.
Green
Heat - Shrink shroud
Load Cell 3.
Red
Heat - Shrink shroud
Load Cell 4.
Yellow
Heat - Shrink shroud
As we are removing Load Cell 1 in this example we will remove the
Blue
heat - shrink shrouded cable from the Molex
connector. (See Below)
Once this is done we can remove the bolts holding the load cell in place. Use a 18mm socket and the 19mm open
ended spanner to do this as shown below.