ArtiCure ASB Посібник із швидкого старту - Сторінка 4

Переглянути онлайн або завантажити pdf Посібник із швидкого старту для Медичне обладнання ArtiCure ASB. ArtiCure ASB 4 сторінки.

ArtiCure ASB Посібник із швидкого старту
Troubleshooting Error Codes
• Error codes appear in the bottom of the ASU screen
• "E" Error codes are recoverable and remain on the screen until
the Footswitch is depressed for next cycle
• "H" Error codes will remain on the screen until issue is corrected
• "P" and "F" error codes are non-recoverable and require a restart
(power cycle) of the ASU. If the problem is not resolved, then unit
must be returned for service.
ERROR CODES
ERROR CODE
EO1, E03 Check Electrodes
(Low Impedance Error)
E02 Close Jaws
(High Impedance Error)
E03 Check Electrodes
*When using Coolrail
Linear Pen
®
(MCR1) only*
HO1 Replace Handpiece
(Invalid Handpiece)
H02 Replace Handpiece
(Time Expired Error)
F13 Power Measurement Error
Code PO1 (Power Generation/
Measurement)
Information relative to other less common error codes can be found in the troubleshooting section in the P000463 ASU IFU.
For other concerns, contact your AtriCure representative or contact
the AtriCure Customer Service.
For detailed information, please consult the Instructions for Use.
4
IFU-0001.C
CAUSE
• Electrodes are shorted (e.g., clamped on
a staple)
• Jaws immersed in a conductive fluid such
as saline or blood pool
• Jaws of clamp not closed on tissue
• Electrical isolative material such as a
suction tube between the jaws
• Broken interconnect cable
• A hi-temp error has occurred indicating
that the end effector has overheated, or
the cooling system has malfunctioned.
• Invalid handpiece
• The 8 hour handpiece clock has expired
• Internal ASU power calculation
• Handpiece electrodes are shorted
Note: Displays the location where an error code will
display. In this example, an E02 code indicated to
close the jaws of the Synergy Ablation Clamp.
REMEDY
• Reposition clamp
• Reposition clamp
• Replace interconnect cable
• Ensure that the fluid cable is not kinked
or occluded and that there is a sufficient
amount of water in the pump box. If these
conditions continue, the Coolrail device
will need to be replaced.
• Replace handpiece
• Replace the handpiece
• Recycle power
• Reposition jaws
• If won't correct, contact customer service
• Recycle power
• If won't correct, contact customer service