gold of the electrode is visible
and coagulum is removed.
8.3. If the Pen is idle between
ablations, place the Pen distal
tip onto saline-soaked gauze to
prevent any coagulum not
cleaned off the electrodes from
drying.
9.
Repeat ablation if necessary.
10. Upon completion of the surgical
procedure, disconnect the Pen from
the ASU/ASB and discard the Pen
after use. Follow local governing
ordinances and recycling plans
regarding disposal or recycling of
device component.
8
INSTRUCTIONS FOR USE
Pacing and Mapping Mode
1.
Inspect the package and product to
ensure the expiration date has not
passed and no damage occurred to
the product during shipping and
handling.
2.
Using sterile technique, remove the
Pen from its packaging.
3.
With the Connector Alignment Arrow
symbol in the 12 o'clock position, push
the Connector into the Pen receptacle
on the front of the ASB. Verify that the
connections between the Pen and the
ASB are secure. If the connections are
loose, do not use the Pen. Inspect the
Cable and do not use the Pen if the
cable is frayed or the insulation is
damaged.
Insert with Connector Alignment
Arrow at the 12 o'clock position
4.
A cable with standard 0.080 inch pin
tip plugs is provided with the ASB to
connect to the auxiliary device. The
end of the cable with the single pin tip
plug is connected to the ASU Source
Switch. The end of the cable with dual
pin tip plugs is connected to the
auxiliary device.
5.
Turn on the temporary pacing or
sensing equipment and ensure proper
connections to validate electrical
continuity. For detailed instructions
refer to temporary pacemaker manual.
6.
Settings and procedures for the
auxiliary device are determined
according to the instructions for use
provided with the auxiliary device. Set
auxiliary device to atrial asynchronous
mode (sensing disabled or increased
to maximum value).
Note: The Pen will pace when the auxiliary
device is in the ON position.