PD030 DIODE LASER MANUAL
4. Operation Instructions continued.

4.5 Automatic Calibration

When the power is set up, user can calibrate the output power:
1.
Set up the power, then insert the fiber pen into the power meter hole, the system
would change to auto calibration interface.
2.
3.
Press OK to start the calibration.
4.
Please operate according to the notice until the calibration is finished, then take out
the fiber pen.
Caution: Do not take out the fiber pen when calibration is in process.
When calibrating, the areas are locked except the notice area.
DLC Australia Pty Ltd 11
www.dlc.com.au