DocuPen Color Calibration procedure
Note:
You need letter-sized printouts for both Color (photo-paper is recom-
mended for Color calibration) and Roller Calibration. Please download the files
from the CD and print them before starting the calibration procedure
The DocuPen is factory calibrated and there should be no need to re-calibrate. We
advise you to recalibrate only if you experience poor scan quality, e.g. scanned image
appears in white or black - meaning that the color conversion tables have been lost.
Note:
Your DocuPen must be
Calibration procedure
4. You will be prompted to turn the DocuPen ON. Press the Power On button on your
DocuPen and Click OK in the Dialog window.
Note:
Your DocuPen memory will be erased!
DocuPen Operation Manual 22
CONNECTED
to your PC before starting the
1. Open the DocuPen Scanner Control Window.
Click on Options and select Calibration.
2. In the Calibration
Dialog window that
appears click on the Auto
Calibration button
3. In the Calibration Setup window that
appears please Calibrate Color and click
OK.
DocuPen Color Calibration Procedure (continued)
After the memory has been erased the Calibration scanning window will appear.
5.
Disconnect
the DocuPen and scan the Color Calibration sheet. For optimal
results you must scan in Photo Mode and in Hi-Rez. The
will start flashing indicating that the DocuPen is in calibration mode. Position the
DocuPen on top of the Color Calibration sheet and glide it over the image from top to
bottom. Stop the movement about 2 cm below the last gray band.
6. The DocuPen will now save the Calibraton image into memory. After the calibration
image is saved into DocuPen's memory,
click OK in the Calibration scanning window. You will be prompted to turn the
DocuPen ON.
Green OK
You should see the
sign in the Calibration result window - You have
successfully calibrated your DocuPen!
7. Finally, click Upload Values to pen button in the Calibration result window to send
the calibration result to the DocuPen. You will be prompted to turn the DocuPen ON.
Turn ON the scanner and click OK in the dialog window.
DocuPen Operation Manual 23
Red
LED on the DocuPen
connect
the DocuPen to your computer and