Etac Molift Air Series Посібник користувача - Сторінка 10

Переглянути онлайн або завантажити pdf Посібник користувача для Допомога з мобільності Etac Molift Air Series. Etac Molift Air Series 15 сторінок.

Etac Molift Air Series Посібник користувача
Molift Air

Transport and Operating Conditions

Transport and Storage
For long time storage it is recommended that the
emergency stop button is activated (pulled out).
The lifter can be stored and transported under
temperatures between -25 - 70 °C.
Operating
The lifter is designed for use at standard room
temperatures (+5 to +40°C).
Air Pressure: 70 - 106 kPa
Relative Humidity: 15 - 93 %
Following storage or transport at other tempera-
tures leave the lifter in a room with a suitable
temperature until it reaches a safe operating
temperature.
Medical electrical equipment
requires special precautions regard-
ing electromagnetic compatibility
(EMC). Portable or mobile radio
communication equipment may
affect the medical electrical equip-
ment, and should be kept minimum
25 cm (10 inches) from the lifters

electronics.

Batteries

Molift Air has a 26.4 V NiMH battery pack. The
battery has a life expectancy of approx. 500
charge cycles.
Used batteries must be disposed
of as special waste according to
local rules and regulations. Do not
dispose in general waste
Batteries must be fully charged and
depleted completely a couple of
times before it achieves full capac-
x2
ity. Also follow this procedure if
the lifter has not been in use for a
long period (3-4 weeks)
Electronics

Battery and Service indicator

The lifter has two indicator lights.
Wrench symbol / Service light
Battery symbol / Battery light
The electrical system has a power save function
which will turn off the electrical system after ten
minutes without activity. All lights will turn off.
The system is activated when pushing one of the
operating buttons
Battery Light
The battery indicator light will illuminate and
make a sound when battery level is low and the
lifter requires charging. When this occurs the lifter
will have sufficient power available for one full lift-
ing cycle with max load. When battery is critically
low it is only possible to lower the lifting belt.
Service Light
The lifters electronics record the loads lifted and
number of lifts. After a certain period of operation
a signal is given to indicate that service is required.
Service light
No light
Green
Yellow
Red
Red + sound

Service scope

Service involves replacing the lifting belt and
inspection/replacement of worn parts. This must
be carried out by authorized personnel.
Service is needed when the service light is red
(calculated 5.000 lifts in weight class 2).
9
Mode
Power saving (Stand by)
Ready for use
Order service
Perform service
Perform service
immediately