akrometrix Table Top Shadow Moire Посібник користувача - Сторінка 11
Переглянути онлайн або завантажити pdf Посібник користувача для Лабораторне обладнання akrometrix Table Top Shadow Moire. akrometrix Table Top Shadow Moire 14 сторінок.
Table Top Shadow Moiré User Manual
Appendix A - Miscellaneous Information
A.1 Equipment Ratings
Supply Voltage
Full Load Amp Rating
Footprint
Weight
A.2 Installation and Assembly
1. The TTSM is designed to be installed atop a table, in a room with a temperature
between 20–26C and humidity between 20-80% non-condensing. The lighting
should not be overly bright, nor so dim that these instructions cannot be read.
2. The TTSM can be lifted and carried by two people, and should be held by the frame
at the base. Place it on the table where it will be installed.
3. Remove any packing materials attached to the equipment.
4. Unscrew the grating support from the motor stage and remove it.
5. Remove the strap from the motor stage (see Figure A.1 and Figure A.2).
6. Screw the grating support back into place.
7. Install the grating on the grating support.
8. Install the mirror on the back of the grating support.
9. Attach one of the included lenses to the camera. Match the lens with the grating
using Table 1.1.
10. Assemble the monitor and place it on the table next to the unit. Place the keyboard
and mouse on the table.
11. Install the computer where it can be attached to both the monitor and the system.
12. Connect all cables.
a. The TTSM unit and monitor must be plugged into wall sockets.
b. The Ethernet cable on the back of the unit connects to the top Ethernet add-on
card port on the back of the computer.
c. The RS2-32 cable on the back of the unit connects to the S1 port on the
computer's break-out serial port cable.
d. Monitor cable, keyboard and mouse cables, Ethernet cable to local network
connects to motherboard Ethernet port.
Electrical
100 – 240 VAC, 50-60 Hz
10 A
Mechanical
490 (w) x 531 (d) x 678 (h) mm
32 kg (70.5 lbs)
40-5000-8100 Rev. A
Page 11 of 14