Ashcroft 2274 XAO Інструкція з експлуатації Посібник з експлуатації - Сторінка 5

Переглянути онлайн або завантажити pdf Інструкція з експлуатації Посібник з експлуатації для Вимірювальні прилади Ashcroft 2274 XAO. Ashcroft 2274 XAO 13 сторінок. Digital industrial gauge

Ashcroft 2274  XAO Інструкція з експлуатації Посібник з експлуатації
KEYPAD FUNCTIONS
This key allows for selecting gauge features in the
menu finalizing selection. Use of the enter key is
ENTER
described throughout operating instructions.
MENU OPTIONS
UNITS: 12 units of measurement are available: psi, mmHG,
inH
O with three temperature options: 20°C, 60°F, 4°C*,
2
mBar, inHg, ftH
O, mPa, kPa, kg/cm
2
Step 1: Press the MENU key until the word UNITS appears.
Step 2: Press Enter.
Step 3: Press the
(up arrow key) or
key) to select the required unit of measure.
Step 4: Press ENTER to finalize the UNIT selection.
Note: For inH
0 range with three temperature options,
2
press the
(up arrow key) and
select the desired temperature, then press ENTER to final-
ize the UNIT selection.
CONFIG: This option allows access to additional Menu
options. Options include:
• ENTPW or enter password (this appears as a sub-menu
in the CONFIG mode if a user password has been set).
• RECAL (allows for zero, span and mid-scale calibra-
tion of the gauge).
• 0bUTN or zero key (allows for adjustment of % of
range that can be zeroed),
• dISAb, allows for disabling MENU options.
• SETPW: This feature allows for a user defined numer-
ic password. If a user password is not set, all features
in the CONFIG mode will be accessible without a
password. If a user password is set, all items in the
CONFIG menu options are accessible with or without
a user password. If a user password is programmed,
it will be required to access the CONFIG menu options.
*Indicates default.
– 8 –
2
and bar.
(down arrow
(down arrow key) to
KEYPAD FUNCTIONS
MENU FUNCTIONS
How to Use Your Menu Functions
To set a user password (SETPW):
Step 1: Press the Menu key on the keypad
Step 2: Press the
(up arrow key) or
key) until the word CONFIG appears.
Step 3: Press Enter. The word SETPW appears on the
gauge display
Step 4: Press Enter. A five digit numeric password is now
required.
Step 5: Press the
(up arrow key) or
key) on the keypad to select the first digit of the password.
Step 6: Press ENTER.
Step 7: Repeat until the five-digit password is shown on
the gauge display.
Step 8: Press ENTER.
Note: to erase password at any time while in the SETPW
(set password) mode, press the ZERO/CLEAR key. The
user will be prompted to reprogram the password once the
five-digit password is entered. The gauge will display SAVE.
Step 9: Press Enter to save the password setting.
ENTPW: Once a user password has been established and
entry into the CONFIG mode is required, the user will be
prompted to ENTPW or enter password.
Follow setup steps 4-8 above.
RECAL: or recalibrate allows for zero, mid-scale, full-scale
and factory default calibration of the gauge. The RECAL
feature also allows for recalibration of gauges with
4-20mA output.
*Indicates default.
– 9 –
(down arrow
(down arrow