EG&G ORTEC 463 Посібник з експлуатації та обслуговування - Сторінка 16
Переглянути онлайн або завантажити pdf Посібник з експлуатації та обслуговування для Вимірювальні прилади EG&G ORTEC 463. EG&G ORTEC 463 20 сторінок. Constant fraction discriminator
10
pulse from the preamplifier fast output will be less than
'V20 nsec. For such systems the 463 should be operated
in the Scint/SB mode for a time to crossover of "^2 nsec.
For surface barrier systems with the ORTEC 454 Timing
Filter Amplifier operating on the preamplifier slow output,
with a rise time of >20 nsec, the 463 should operate in the
Ge(Li) mode. The time to crossover in the Ge(Li) mode
is 'V/IS nsec; so this shaping is best for signals that originate
in most charge-sensitive preamplifiers.
Figure 7.8 is a typical timing spectrum taken with the 463
and using the system shown in Fig. 7.7 for a detector of
100-/L(m depletion depth. The timing spectrum in Fig. 7.9
was obtained with the same system for a detector with a
depletion depth of 1975 /jm. These two spectra illustrate
the usefulness of the surface barrier detector system in
a pulsed distance-measuring application. The distance reso
lution for a single laser burst with a light return equivalent
to 5.5 MeV would be about 3 in., and this intensity is
smal l compared to the output capability of such lasers.
If the time interval is averaged for several laser bursts, the
distance resolution could be reduced to only a smal l
fraction of an inch.
/\
Detector:
•
•
Area • 100 mm2
•
Depth • 100 pm
,
463 Discriminator
Alpha Equivalent Energy of 5.48 MeV
^
•
2 nsec
•
FWHM = 255 psec
•
<
\
/
••
..
N-v
Channel Number
SB Timing Detector SN 10-4760
Area; 50 mm^. Bias = 400V
Depth: 1975Aim
Alpha Energy Resolution = 29.6 keV
^
RCA Laser Equivalent Light =
y-y
5.5 MeV
• /*.
Start Channel: Laser Pulser
/•**• i
Stop Channel: 454,463
••
Calibration: 8.7 psec/Channel
Gamma Rise Time = 15 nsec
FWHM'-' 660 psec \
y
\
.•vC
Fig. 7.8. Typical Timing Spectrum for a 100-/im Surface
Barrier Detector Using the System of Fig. 7.7.
Fig. 7.9. Typical Timing Spectrum for a 1950-/im Surface
Barrier Detector Using the System of Fig. 7.7.
7.4. TIMING WITH HIGH-RESOLUTION
LOW-ENERGY PHOTON SPECTROMETERS
A system that i llustrates a method for measuring l ifetimes
of excited states at extremely low energies is shown in
Fig. 7.10. The start channel for the ORTEC 457 Time to
Pulse Fleight Converter is a plastic scintillation detector
with an RCA—8575 photomultiplier tube and uses the 463
set for the Scint/SB
mode as the fast discriminator.
The stop channel is a 6-mm Low Energy Photon Spectrom
eter (LEPS) Ge(Li) detector operating with a fast timing
system composed of a 454 Timing Filter Amplifier and a
463 Constant Fraction Discriminator. The 454 is operated
with its Int (Integrate) not greater than 10 nsec and the
□iff (Differentiate) >100 nsec. The 454 gain should be
adjusted so that the highest energy of interest corresponds
to 'V'lO V at the 454 output. Operate the 463 in its
Ge(Li)
mode, with time to crossover about
18 nsec.
Figure 7.11 shows the timing resolution versus deposited
energy for a typical 6-mm LEPS detector. The system of
Fig. 7.10 was used to obtain the data points. A ^^Na
source is used to produce two coincident 511-keV gamma
rays from positron annihilation. The start channel detects
one of the gammas, and a Compton-scattered event in the
LEPS wil l occasional ly produce the energy selected by the
451, 455, and 416 slow channel to gate the multichannel
analyzer and permit time to be measured for that coin
cident event. By selecting the energy of interest with the