Make sure that the DRAIN Valve is closed (horizontal). Fill the BATH with tap water to the bottom of the
SHELF if you have one CORRAL mounted and to the screw of the shelf if you have 2 CORRALs mounted.
If your tap water contains a lot of lime, use distilled or de-ionized water, add a pinch (0.01gr.) of salt to make
the water conductive so that the level sensors will work. Do NOT use saline solution. . Add Algaecide Part
#ALG240 (1.0 ml. per 4 liter of water. Available USA only) to the water. Any other bath clearing agent can
also be used.
Make sure the power switch is OFF. Plug the power cord into a properly grounded electrical outlet.
Turn the power switch ON. The pump will start pumping the water and the display will indicate the temperature.
The unit will heat up to 37.0
Use the LID over the bath to prevent evaporation and help keep the water cleaner. DO NOT USE THE LID
WHEN THAWING, as water condensate from the lid may drip onto the plasma bag ports.
Open the LID by lifting from the front. Hold the front of the LID up to let the condensate flow back into the Bath,
then lift up the LID and place it aside.
o
C. The display will show "HEAT ON" when the unit is heating.
3