Baumer FLDK 110G1303/S14 Benutzerhandbuch - Seite 14

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Kassenschalter Baumer FLDK 110G1303/S14 herunter. Baumer FLDK 110G1303/S14 35 Seiten. Laser copy counter

Baumer FLDK 110G1303/S14 Benutzerhandbuch
8
Adjustments
With the SCATEC-2, the following parameters and operating modes can be adjusted:
Output pulse length
False pulse suppression
Sensitivity
The parameters can be adjusted in two ways: via the interface or using DIP switches.
Parameter setting through the interface allows continuous choice of the parameters, whereas a selection
is made from a defined set when the parameters are set with the DIP switches. Parameter setting through
the interface using a computer is described in the user manual for the software package ScaDiag.
Note:
The sensor is only in DIP switch mode when the power LED lights green.
If the power LED lights yellow instead of green, the parameters of the sensor have been set through
the interface. In this case, the DIP switches are inactive on the sensor and their settings are
irrelevant. The sensor can be reset to DIP switch mode, in which the parameters are defined by the
DIP switches, in two ways.
a)
Through the interface (see user manual for the software package ScaDiag).
b)
By setting the DIP switches in the following sequence:
1.
Set all switches to OFF (at least one must have been ON previously)
2.
Set all switches to ON within 16 seconds
3.
Set all switches back to OFF within 16 seconds
4.
The power LED should then light green after a brief delay. The parameters of the
sensor can then be adjusted with the DIP switches
To prevent the loss of the DIP switch cover screw, do not fully remove it from the cover. The screw
is retained by the cover.
Manual SCATEC-2
Version 2021-09
14
www.baumer.com
Baumer Electric AG
Frauenfeld, Switzerland