I
S
NITIAL
ETUP
If any problems are found during the initial setup procedure, refer to the
troubleshooting section on page 10. For further assistance, contact
Drucker Diagnostics at 814-692-7661.
1. Plug the centrifuge in to an approved electrical outlet. For
electrical safety, the unit must always be properly grounded.
2. Spin the rotor by hand, check for free and level rotation. If the
rotor does not spin freely refer to page 10 on troubleshooting.
3. Place the six test tube holders inside the rotor and verify that
they are seated properly.
4. Close the lid. Rotate the lid knob clockwise to its complete stop
position.
5. Turn the centrifuge on by turning the timer to 10 minutes.
6. The test tube holders will slide up into the horizontal position
and the unit will accelerate to full speed.
7. Listen to the sound of the centrifuge. A smooth whirring sound
should be heard. If there are any loud or unusual sounds, stop
the centrifuge by turning the timer to zero (0).
8. While the machine is running, try to turn the latch
counterclockwise.
9. Power should be cut to the motor
10. If the motor power is not cut when the lid latch is opened,
contact Drucker Diagnostics for assistance
After the centrifuge has passed this procedure, it is ready for operation.
Page | 5 Drucker Diagnostics +1-814-692-7761 - [email protected]