Anyload EC200 Benutzerhandbuch - Seite 8

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Skalen Anyload EC200 herunter. Anyload EC200 8 Seiten.
Auch für Anyload EC200: Betriebshandbuch (16 seiten)

Anyload EC200 Benutzerhandbuch
7.3 Scale 3 Calibration:
Connect the extra platform through DB9.
SCALE3 DB9
1
S-
2
S+
3/6 V+
4
V-
5
GND
7~9 NC
Turn on the scale, and then press <1>, <3>, <1>, <3>, <Enter> when the scale
examining itself.
7.3.1 Menu setting:
'WEIGHT display' shows
'UNIT WEIGHT display' shows CAP XXXX.
'TOTAL display' will show d=X.X.
Then press <UNIT> button to select calibration unit between kg and lb,Press <CE>
button and then press 'number buttons' to set the max calibration value,press<*>
key to select division,press <Enter> key to confirm。
7.3.2 Calibration:
Make sure nothing loaded on the external platform.
Press <Enter> to set zero point. (UNIT WEIGHT display will show CAL-0 flashing
for a few seconds and then changes to XXXX to set the Calibration weight).
Press <CE> button and then press <NUMERIC (0-9) KEY> to set the calibration
weight. Calibration weight will show on the UNIT WEIGHT display. Put the test
weight on the external platform. Press <ENTER> button, the UNIT WEIGHT display
will show a flashing XX then change to XX.XXX lb. Then the Calibration procedure
complete.
7.3.3 Check Calibration:
Reboot the scale. Place test weight on the scale and see the weight is correct or
not. If not, please redo the above Calibration procedure.
Page 8 of 8
SCALE3,
Load cell DB9
1
S-
2
S+
3
V+
4
V-
5
GND
6~9 NC
AL-Manual-EC200-En-101