Warning Messages
The following warning messages are displayed at power up, if relevant. Any key press
clears the message.
The following error messages may be displayed for other faults during normal sampling
applications:-
The Dashboard Display Screens
Blockage Retry: If the pump cannot maintain the target
flow rate within 5% for more than 20 seconds (due to
say a kinked tube or inlet blockage) then it will
automatically halt sampling. After one minute the pump
will attempt to re-start. If the specified number of
attempts to retry is unsuccessful, the pump will
terminate the current sampling run.
The Run or Stop Mode 'dashboard' displays show data
from either the previous or the current run. The screen
title bars are colour coded; red for 'stopped' and green
for when the pump is running. They detail:
• Battery Life, either % or estimated hours (in Plus &
Pro)
• Back Pressure
• Motion %
• Flow rate
• Volume
• Elapsed Duration of run