4. After the wicking sheet is in position, place the tray insert into a UVT tray at 90 degrees to each
other so that the gel is centered between, but not touching, the rounded edges of the insert
tray. Again, be careful not to trap air pockets between the tray insert and the wicking sheet
(Figure 3). With a gel plate just center the gel on the wicking sheet and insert the tray.
5. Hold the UVT tray (or gel plate) and tray/insert assembly securely but gently together and invert
the entire assembly.
6. Carefully remove the UVT tray or gel plate, leaving the gel in place on the wicking sheet on the
now inverted insert tray (Figure 4).
Note: The wicking sheet runs over the insert tray legs.
Figure 3: Gel and wicking sheet on tray insert
Figure 4: Inverted insert tray