Datexel DAT8016 Setup-Handbuch - Seite 2
Blättern Sie online oder laden Sie pdf Setup-Handbuch für E/A-Systeme Datexel DAT8016 herunter. Datexel DAT8016 4 Seiten. Modbus tcp th ermocouple input module
Select the Language.
1. Click English.
2. Click enter.
Th e device name should appear in the center
of the page. You are also able to "Logout" of the
DAT8016, go "home", or go "back" to the page
before, by clicking the icons on the page.
How to read the Analog Input
Signal.
1. Click the Analog Input at the bottom of the
page.
2. Select the type of input being measured.
3. Select Tc K for K type thermocouple.
4. Click Save to save to the DAT8016.
5. Click Read to measure the Input.
6. Adjust the calibrator to check the range.
Th e input selection can be seen below.
250mV, J type thermocouples, K type ther-
mocouples, R type thermocouples, S type
thermocouples, T type thermocouples, B type
thermocouples, E type thermocouples, N type
thermocouples.
Each input can be diff erent and they are all iso-
lated from each other.
Th e red indication under Break means the input
is open circuit. By clicking the "Set Same Input"
button you enable all the inputs to be the same as
Analog Input 0.
2 of 4
2 of 4