6. Press and release the Light Source Key to select the light source being measured. The different light
source options are:
Flood Lamp - For use when measuring energy of a flood-lamp light source.
3-mm Lightguide - For use when measuring energy at the end of a 3-mm lightguide.
5-mm Lightguide - For use when measuring energy at the end of a 5-mm lightguide.
8-mm Lightguide - For use when measuring energy at the end of a 8-mm lightguide.
7. Press and release the Mode Key to select between the meters operating modes. The different
operating modes are:
Peak Intensity - Shows the highest visible-light intensity in mW/cm
the measurement.
Intensity - Shows the visible-light intensity in mW/cm
Dose - Shows the total visible-light dose in mJ/cm

Maintenance

The ACCU-CAL™ 50V was designed to operate with minimum maintenance. Follow the schedule below to
assure top unit performance.
Calibrate the instrument at least annually. Calibration service is available through Dymax Customer
Service or Dymax Product Repair.
Change the batteries when a low battery warning is received. The ACCU-CAL™ 50V uses two AA-type
batteries. The battery compartment is on the back of the instrument.
Figure 8. Battery Compartment (Closed & Open)
Keep the detector head's sensing element clean and free of contaminants. The detector head may be
cleaned with a clean tissue wetted with isopropyl alcohol.
Dymax ACCU-CAL™ 50V Radiometer User Guide