dymax Accu-Cal 50 Руководство пользователя - Страница 9
Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы dymax Accu-Cal 50. dymax Accu-Cal 50 12 страниц. Radiometer
Также для dymax Accu-Cal 50: Руководство пользователя (12 страниц), Руководство пользователя (14 страниц)
6. Press and release the Light Source Key to select the light source being measured. The different light
source options are:
Flood Lamp - For use when measuring energy of a flood-lamp light source.
3-mm Lightguide - For use when measuring energy at the end of a 3-mm lightguide.
5-mm Lightguide - For use when measuring energy at the end of a 5-mm lightguide.
8-mm Lightguide - For use when measuring energy at the end of a 8-mm lightguide.
7. Press and release the Mode Key to select between the meters operating modes. The different
operating modes are:
Peak Intensity - Shows the highest visible-light intensity in mW/cm
the measurement.
Intensity - Shows the visible-light intensity in mW/cm
Dose - Shows the total visible-light dose in mJ/cm
Maintenance
The ACCU-CAL™ 50V was designed to operate with minimum maintenance. Follow the schedule below to
assure top unit performance.
•
Calibrate the instrument at least annually. Calibration service is available through Dymax Customer
Service or Dymax Product Repair.
•
Change the batteries when a low battery warning is received. The ACCU-CAL™ 50V uses two AA-type
batteries. The battery compartment is on the back of the instrument.
Figure 8. Battery Compartment (Closed & Open)
•
Keep the detector head's sensing element clean and free of contaminants. The detector head may be
cleaned with a clean tissue wetted with isopropyl alcohol.
Dymax ACCU-CAL™ 50V Radiometer User Guide
2
seen by the detector during
2
at the detector during the measurement.
2
at the detector during the measurement.
9