Scanning your 404 LE

Always place the phantom on a stable, level surface for scanning.
The phantom comes ready to scan. Do not peel off the surface material.
Do not press the transducer into the scanning surface. This damages the
scanning surface and will shorten the life of the phantom. For curved transducers, use
water or a layer of gel.
Use water or a generous amount of coupling gel to ensure good transmission. Do
not use mineral oil, baby oil or lanolin-based gels as a coupling medium. Poor transmis-
sion is a result of insufficient coupling.
Clean the scanning surface immediately after use. Use a soft cloth or paper towel
and soap and water, if needed.
Caution:
Do not press the transducer
into the scanning surface.