Scanning your 404 LE
•
Always place the phantom on a stable, level surface for scanning.
•
The phantom comes ready to scan. Do not peel off the surface material.
•
Do not press the transducer into the scanning surface. This damages the
scanning surface and will shorten the life of the phantom. For curved transducers, use
water or a layer of gel.
•
Use water or a generous amount of coupling gel to ensure good transmission. Do
not use mineral oil, baby oil or lanolin-based gels as a coupling medium. Poor transmis-
sion is a result of insufficient coupling.
•
Clean the scanning surface immediately after use. Use a soft cloth or paper towel
and soap and water, if needed.
Caution:
Do not press the transducer
into the scanning surface.
5