Fault codes
Prime and Dual models only. Ambient
Fault codes are shown in the Speed section of the LED display (shown on previous page
The following List show what each Code means and what to do.
F A 5 0
F A 2 1
Non Fault Codes
Speed will not go beyond 2000 Rpm.
.
The Electronics have set the Speed to safe mode
.
Check the rotor bottom has magnets fitted
.
Check correct rotor is in use against program chosen
.
If not then
Noisy
.
Unbalanced tubes, see Rotor pages
.
Lid seal needs replacing
.
Rotor Lid not fitted
Lid support fail
.
Spring has failed.
Inverter Drive
Inverter error
*
Contact our Service Department
Contact our Service Department
Look at the Obround viewing port (Rear)
Note the Code shown in the inverter display.
Turn off the power at the wall mains switch only.
WAIT 3 minutes, Turn on again and Run. If problem persists
Contact our Service Department.
As above.