a&p instruments 2001 Series Gebrauchsanweisung - Seite 5

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Prüfgeräte a&p instruments 2001 Series herunter. a&p instruments 2001 Series 10 Seiten.

4. Parameter setting
Different parameters can be adjusted by using the Tare together with the Change
button. This is only possible with the radiator in the backwards position and the
display showing the weight.
Push the Tare button very short will tare the balance to zero, push it some seconds
will start the parameter setting mode. Each push on the Tare button now will take you
to the next parameter. Changes are done by using the Change button. This mode
can be stopped at any time by using the SET button.
If you have reached the settings level accidentally, press the SET button and you are
back in the weighing mode.
DISPLAY
shows
Press
„Tara"
PNr.
button
for two
seconds
IA
FEUCHTE
(HU)
Print
K
O
N
GAb
S
T
AIn
A
Abn
S
T
PFl
O
P
2
Heat
Delay

Settings level

SETTINGS can be changed by
using the „Change" button
0000; 1000, ...9000 (Standard is no
sample no.)
0"6"12"30"
Sek.1'2'4' min.
Atro (AO); Dry Matter (dM); Per
1000(PM)
0; 1; 2; 3 prints
4; 8; 16; 32 mg
10", 20"; 45"; 70" sec
0,3'; 1'; 2'; 3' min
aus, 0, ....100% (dM)
0,
....100% Power
75 %
0, .... 9 sec.
MEANING
Sample no., (0001,0002,...)
Printing in intervals
Results
Number of prints
To send data via the interface, „Print"
has to be set at least to „1" !
Weight loss
Weight loss in how many seconds
Start of detecting weight loss
Preflash
for very wet samples to shorten drying
time !
Power of radiator:
with 250W radiator up to app. 160°C
UX 2011 Q up to app. 360°C
Time delay, until the balance´ display
will be steady !
1
3
5
S
T
O
P
B
Y
S
E
T
B
U
T
T
O
N