I n s t r u c ti o n s for No r m a l O p t ic a l Powe r M e te r Fu n c t i o n
6 .2 .
6.2 . 1 .
Press
button to open the instrument, use
the Optical Power Meter measurement, then press
function.
or
measurement for example:
6. 2 .2 .
Co nn e c t w it h t h e tes te d o p tic f ib e r
Clean the optic Fiber end-face, and connect with the PIN port on the instrument (SC/
FC/ST adaptor), noted with the type of the optic fiber adaptor, the non-matched
connecting will make the damage of the connector on the instrument and also make
the faulty of the measurement result.
E.g.: 1310nm measurement, after connect with the optic fiber, it will displaying
-20dBm, press
, set the -20dBm as the reference value, and it will also display
relative value (dB), see as following:
In order to get the loss between the two measurements, do another time
testing under the first relative output power. The relative value will be calculated
automatically. For example the second time testing of 1310nm output power is
-35dBm, the relative output power will be -15dBm. (relative output power=absolute
output power-reference value -15dBm=-35.00dBm- (-20.00dBm) see as following: