Olympus BX61 Anleitung Handbuch - Seite 28

Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus BX61 herunter. Olympus BX61 40 Seiten. Motorized system microscopes

Olympus BX61 Anleitung Handbuch
Problem
c) Coarse adjustment will not go all
the way down.
d) Objective makes contact with speci-
men before focus is obtained.
4. Observation Tube
a) Field of view of one eye does not
match that of the other.
5. Stage
a) Image shifts when you touch stage. Stage is not properly mounted.
b) Specimen stops midway on the X-
axis traverse.
c) X-axis and/or Y-axis knobs are too
tight or too loose.
d) Stroke is decreased.
24
Cause
Condenser holder is too low.
Specimen is mounted upside down.
Interpupillary distance is incorrect.
Incorrect diopter adjustment.
Different eyepieces are used on left and
right.
Your view is not accustomed to micro-
scope observation.
Specimen is not correctly positioned.
X-axis and/or Y-axis adjustment knobs
are set at a position where tension is too
high or too low.
Stage guide is deviated.
Remedy
Raise condenser holder.
Mount specimen correctly.
Adjust interpupillary distance.
Adjust diopter.
Change on eyepiece to match the
other so that both sides are the same
type.
Upon looking into eyepieces, try look-
ing at overall field before concentrat-
ing on specimen range. You may also
find it helpful to look up and into dis-
tance for a moment before looking
back into microscope.
Clamp stage.
Place specimen correctly.
Adjust tension.
Correct deviation as instructed on
page 14.
Page
32
16
16
13
14
14