PHASE CONTRAST OBSERVATION

The following three units are available for phase contrast observation. Mount a slider onto the microscope and replace the
objectives with phase-contrast-compatible objectives.
Model name
Ph precentering slider
CK40-SLP
Ph centering slider
CK40-SL
RP slider
CK40-RPSL
# During relief phase contrast observation using the RP slider, part of the field of view may get dark when the lamp
voltage is lowered. This happens because the viewing field´s brightness is affected by fluorescent light from the
ceiling. Should this happen, turn off the room light or raise the lamp voltage of the unit. When a 10X objective is used,
the edges of the field of view may become dark.
19
Description
· The light annuli are precentered, so no adjust-
ment is required. (For 4X, for 10X/20X/40X &
empty)
· The empty opening can be used as a filter holder.
· The light annulus has to be centered.
(For 4X, for 10X/20X & empty)
· The empty opening can accommodate the CK2-
RS40 light annulus for the 40X objective or can
be used as a filter holder.
· The relief slit has to be centered.
(For 4X, for 10X/20X & empty)
· The empty opening can accommodate the
CK40-RPS40 relief annuls for the 40X objective.
Compatible objectives
SPlan4XPL-6
DAch10XPL-6
LWDCDAch20XPL-6
LWDCDPlan40XFPL-6
SPlan4XPL-6
DAch10XPL-6
LWDCDAch20XPL-6
LWDCDPlan40XPL-6
SPlan4XRP
DAch10XRP
LWDCDAch20XRP
LWDCDPlan40XRP