Olympus CK30 Manual de instrucciones - Página 24

Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus CK30. Olympus CK30 39 páginas. Culture microscope

Olympus CK30 Manual de instrucciones

PHASE CONTRAST OBSERVATION

The following three units are available for phase contrast observation. Mount a slider onto the microscope and replace the
objectives with phase-contrast-compatible objectives.
Model name
Ph precentering slider
CK40-SLP
Ph centering slider
CK40-SL
RP slider
CK40-RPSL
# During relief phase contrast observation using the RP slider, part of the field of view may get dark when the lamp
voltage is lowered. This happens because the viewing field´s brightness is affected by fluorescent light from the
ceiling. Should this happen, turn off the room light or raise the lamp voltage of the unit. When a 10X objective is used,
the edges of the field of view may become dark.
19
Description
· The light annuli are precentered, so no adjust-
ment is required. (For 4X, for 10X/20X/40X &
empty)
· The empty opening can be used as a filter holder.
· The light annulus has to be centered.
(For 4X, for 10X/20X & empty)
· The empty opening can accommodate the CK2-
RS40 light annulus for the 40X objective or can
be used as a filter holder.
· The relief slit has to be centered.
(For 4X, for 10X/20X & empty)
· The empty opening can accommodate the
CK40-RPS40 relief annuls for the 40X objective.
Compatible objectives
SPlan4XPL-6
DAch10XPL-6
LWDCDAch20XPL-6
LWDCDPlan40XFPL-6
SPlan4XPL-6
DAch10XPL-6
LWDCDAch20XPL-6
LWDCDPlan40XPL-6
SPlan4XRP
DAch10XRP
LWDCDAch20XRP
LWDCDPlan40XRP