SPECIFICATIONS
Item
Motorized Reflected
Fluorescence Illuminator
BX-RFAA
Mercury lamp housing
Operating environment
22
· UIS (Universal Infinity System) optics
· Magnification: 1X (Superwide field compatible: FN 26.5)
· Observation switching: Motorized mirror unit turret (Switching rate approx. 0.8 sec.)
Mirror unit turret carrying max. 6 mirror units.
· Aperture and field iris diaphragms: both centrable and detachable
· Shutter provided: Motorized opening/closing), release speed 0.1 sec.
· Motorized revolving nosepiece connector: Connector position fixed at the factory.
· Slider insertion slots:
@Analyzer/6-position barrier filter slider
²6-position filter slider
³ND filter slider
|6-position filter slider
· Available observations:
@Reflected fluorescence
²Reflected fluorescence + Transmitted light DIC
³Reflected fluorescence + Phase contrast
|Transmitted light
· Applicable microscope frames: BX41, BX51, BX52, BX61, BX62, BX51WI, BX62WI
· Options
@Exciter balancer U-EXBABG, U-EXBAUB, U-EXBAUG
²Rectangular field stop BX-RFSS
· 100 W mercury lamp housing U-LH100HG
· 100 W mercury apo lamp housing U-LH100HGAPO
· Mercury burner USH102D (USHIO) or HBO103W/2 (OSRAM)
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly
through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations: ±10%
· Pollution degree: 2 (in accordance with IEC664)
· Installation (overvoltage) category: II (in accordance with IEC664)
Specification