Attaching the Fluorescence Mirror Units
@
1. Using the Allen screwdriver, loosen the clamping screw @ at the right
side of the illuminator.
2. Pull out the turret and place it upside down.
}Dummy mirror units ² are mounted in the mirror unit positions. Remove
the dummy mirror units from the positions you want to mount mirror units
by loosening the clamping screw ³ of each mirror unit position using
the Allen screwdriver.
|
3. Hold the fluorescence mirror unit | to be mounted so that the model
name inscription on the side is upside down, align it with the mount
dovetail and insert all the way into the insertion position. Tighten the
clamping screw ³ firmly.
# If the clamping screw ³ is loose, the turret will be unable to be
rotated due to interference with the turret cover.
4. Check the mount dovetail number ƒ and place upside down the
indication sheet of the mounted fluorescence mirror unit into the
indication pocket ... with the same number on the front of the turret.
5. Mount other required fluorescence mirror units in the visible mount dove-
tails by repeating the above steps for each of them.
}If you want to mount more fluorescence mirror units in the mount dove-
tails located under the turret cover, first complete the present procedure
then rotate the turret by motorized operation then mount the mirror units
in the mount dovetails.
6. Place the turret in the original position and tighten the clamping screw @
while pushing the turret in.
(Figs. 11 & 12)