AccuMed CMS-50D Benutzerhandbuch - Seite 2
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Medizinische Ausrüstung AccuMed CMS-50D herunter. AccuMed CMS-50D 2 Seiten. Fingertip pulse oximeter
Auch für AccuMed CMS-50D: Handbuch (14 seiten)
4 Technical Specifications
1) Display Format: OLED Display;
SpO
Measuring Range: 0% - 100%;
2
Pulse Rate Measuring Range: 30 bpm - 250 bpm;
Pulse Wave Display: columniation display and the waveform display.
2) Power Requirements: 2 ×1.5V AAA alkaline batteries (or using the rechargeable battery instead),
adaptable range: 2.6V~3.6V.
3) Power Consumption: Lower than 30mA.
4) Resolution: 1% for SpO
and 1 bpm for Pulse Rate.
2
5) Measurement Accuracy: ±2% in stage of 70%-100% SpO
smaller
than 70%. ±2 bpm or±2% (select larger) for Pulse Rate.
6) Measurement Performance in Weak Filling Condition: SpO
correctly when pulse-filling ratio is 0.4%. SpO
larger).
7) Resistance to surrounding light: The deviation between the value measured in the condition of
man-made light or indoor natural light and that of darkroom is less than ±1%.
8) It is equipped with a function switch. The Oximeter can be powered off in case there is no finger in
the Oximeter for 5 seconds.
9) Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
5 Accessories
One hanging rope;
Two batteries (optional);
One User Manual.
6 Installation
6.1 View of the Front Panel
Figure 2 Front view
6.2 Battery
Step 1. Refer to Figure 3 and insert the two AAA size batteries in the right direction.
Step 2. Replace the cover.
Please take care when you insert the batteries as improper insertion may damage the device.
Figure 3 Batteries installation
6.3 Mounting the Hanging Rope
Step 1. Put the end of the rope through the hole.
Step 2. Put another end of the rope through the first one and then tighten it.
Figure 4 Mounting the hanging rope
7 Operating Guide
1) Insert the two batteries in the correct direction, and then replace the cover.
2) Open the clip as shown in Figure 5.
Figure 5 Put finger in position
3) Put the patient's finger into the rubber cushions of the clip (make sure the finger is in the right
position), and then clip the finger.
4) Press the button once on front panel.
5) Do not shake the finger and keep the patient at ease during the process. It is not recommended for
the patient to be moving.
6) Get the information directly from screen display.
7) The button
has three functions. When the device is power off, pressing the button can turn it
on; when the device is powered on, pressing the button shortly can change direction of the screen;
when the device is powered on, long pressing the button can change brightness of the screen.
Fingernail and the screen should be on the same side.
8 Repairing and Maintenance
Please change the batteries when the low-voltage displayed on the screen.
Please clean the surface of the device before using. Wipe the device with medical alcohol first, and
then let it dry in air or clean it with dry clean fabric.
Use the medical alcohol to disinfect the product after use prevents cross infection the next time it is
, and meaningless when stage being
2
and pulse rate can be shown
2
error is ±4%, pulse rate error is ±2 bpm or ±2% (select
2
used.
Please take out the batteries if the oximeter is not in use for a long period of time.
The best storage environment of the device is - 40ºC to 60ºC ambient temperature and no higher
than 95% relative humidity.
Users are advised to calibrate the device periodically (or according to the calibrating program of
hospital). It also can be performed by the state-appointed agent or just contact us for calibration.
High-pressure sterilization cannot be used on the device.
Do not immerse the device in liquid.
It is recommended that the device should be kept in a dry environment. Humidity may reduce
the life span of the device, or even damage it.
9 Troubleshooting
Trouble
Possible Reason
The SpO
and
1. The finger is not properly
2
Pulse Rate cannot
positioned.
2. The patient's SpO
be displayed
normally
be detected.
1. The finger is not placed inside
The SpO
and
2
deep enough.
Pulse Rate are not
2. The finger is shaking or the
displayed stably
patient is moving.
1. The batteries are dead or
almost dead.
The device cannot
2. The batteries are not inserted
be turned on
properly.
3. The device is defective.
1. The device will power off
The display is off
automatically when it gets no
suddenly
signal within 5 seconds.
2. The batteries are almost dead.
10 Key of Symbols
Symbol
Type BF
Warning – See User Manual
The pulse oxygen saturation (%)
PRbpm
Pulse rate (bpm)
The battery voltage is insufficient (change the battery in time to avoid
inexact measurements)
1. No finger inserted
2. An indicator of signal inadequacy
Battery positive electrode
Battery cathode
1. Power switch
2. Change direction of the screen
3. Change brightness of the screen
SN
Serial number
Alarm inhibit
WEEE (2002/96/EC)
IP22
International Protection
11 Function Specification
Display Information
The Pulse Oxygen Saturation(SpO
)
2
Pulse Rate(PR)
Pulse Intensity (bar-graph)
Pulse wave
SpO
Parameter Specification
2
Measuring range
0%~100%, (the resolution is 1%).
Accuracy
70%~100%:±2%, Below 70% unspecified.
Optical Sensor
Red light (wavelength is 660nm)
Infrared (wavelength is 880nm)
Pulse Parameter Specification
Measuring range
30bpm~250bpm (the resolution is 1 bpm)
Accuracy
±2bpm or±2% select larger
Pulse Intensity
Range
Continuous bar-graph display, the higher display indicates a stronger pulse.
Battery Requirement
1.5V (AAA size) alkaline batteries × 2 or rechargeable batteries
Battery Useful Life
Two batteries can work continually for 20 hours
Dimensions and Weight
Dimensions
Weight
II
Solution
1. Place the finger properly and try
again.
2. Try again; Go to a hospital for a
is too low to
2
diagnosis if you are sure the device
works correctly.
1. Place the finger properly and try
again.
2. Keep the patient calm and at rest.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the local service
center.
1. Normal.
2. Change batteries.
Description
Display Mode
OLED
OLED
OLED bar-graph display
OLED
57(L) × 31(W) × 32(H) mm
About 50g (with the batteries)