DigiVac Bullseye Precision Gauge Piezo Benutzerhandbuch - Seite 12

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Messgeräte DigiVac Bullseye Precision Gauge Piezo herunter. DigiVac Bullseye Precision Gauge Piezo 20 Seiten.

12
In many cases, a gauge sensor may become fouled with oil or other foreign matter. It is often
possible to restore the functionality of contaminated probes with cleaning. If the contaminant
is known, the tube should be filled with a fluid that is known to be a solvent to that
contaminant. As an example, ether is often effective in removing residues of some oils.
Commercial carburetor cleaners are very powerful solvents and are highly effective against
some contaminants.
After cleaning with solvents, the gauge sensor should be completely dried or flushed with a
If this is not done,
volatile solvent to assure that it is dry prior to reinstalling it.
contamination of the system may result.

Bluetooth Operation

The Bullseye Piezo comes equipped with a Bluetooth module, which interfaces with the
Vacuum Gauge Application for remote, real-time vacuum monitoring via Apple Store at
https://appsto.re/us/B1Ix5.i
,or Google Play Store at
https://play.google.com/store/apps/details?id=com.digivac.bullseye
ranging up to 300 feet
under ideal conditions.
To enable Bluetooth monitoring, scroll to the menu section marked Bluetooth, and enable or
disable the signal with the up or down arrow. See the included Vacuum Gauge Application
instructional addendum to enable app functioning.
The 775i is designed for trouble free use with little or no maintenance required. A few best
practices should be followed. While not required for accuracy, it is best to mount sensor in
stem down orientation where the threads are down towards the ground, and the wire is up.
This will facilitate draining of any foreign matter from the sensor to help mitigate physical
obstruction of the diaphragm. Keep the measurement unit clean and dry to prevent foreign