2. Select the calibration method AIR or OXYGEN.
3. Allow the calibration gas to flow into the analyzer and wait for
the sensor voltage to stabilize.
4. Press CONFIRM.
NOTE: Ensure that the sensor voltage value is within the sensor
manufacturer's parameters when exposed to the gas. DIVESOFT
oxygen sensors have an air voltage range of 7 - 13 mV. It is
advisable to verify that the sensor shows linear measurement